Dental Panoramic and Cephalometric Systems
Specifications
| Rotograph Plus | Rotograph Digital | ||
| Exposure Times | Panoramic |
17 s adult, 14 s child
|
|
| TMJ |
10.4 s adult, 9.4 s child
|
||
| Cephalometric |
0.2-3 s in 19 steps
|
N/A
|
|
| Exposure interval |
4 min. (1:16 duty cycle)
|
||
| High voltage |
60-85 kVp, in 5 kV steps
|
||
| Anode current |
10 mA
|
||
| Focal spot |
0.5 mm (IEC336)
|
||
| Power supply voltage |
120 VAC ±10%, single phase, 60 Hz
|
||
| Current load |
10 A max
|
||
|
Power rating
|
1.15 kVA
|
||
| Cassette size | Panoramic, TMJ |
5 x 12 (12.7 x 30 cm), curved
|
N/A
|
| Cephalometric |
8 x 10 (20.3 x 25.4 cm), flat
|
N/A
|
|
| Height if irradiated area on sensor |
N/A
|
5 5/8 (143 mm)
|
|
| Average image magnification |
N/A
|
1.2:1
|
|
| Image matrix |
N/A
|
1536 x 2805 pixels in binning modality
|
|
| CCD sensor resolution |
N/A
|
10.4 lp/mm (pixel size: 48µm)
|
|
| Gray levels |
N/A
|
4096 (12 bits)
|
|
| Required operating system |
N/A
|
Windows XP/2000
|
|
| PC connection |
N/A
|
USB 2.0
|
|
| Cooling |
Ambient
|
||
| Leakage radiation at 1 m |
<0.25 mGy/h
|
||
| Degree of protection |
IP 20
|
||
| Maximum operating temperature range |
+50 °F - +104 °F (+10 °C - +40 °C)
|
||
| Weight |
w/ceph arm: 573 lbs. (260 kg)
w/o ceph arm: 520 lbs. (236 kg) |
540 lbs. (245 kg)
|
|
Strato Specifications
| Strato Digital |
Strato 2000 Film
|
||
|
Essential Features
|
|||
|
Motorized column
|
•
|
•
|
|
|
3-line laser patient positioning
|
•
|
•
|
|
|
DC high voltage supply (high frequency power converter)
|
•
|
•
|
|
|
Microcomputer-controlled movements for multiple projection programs
|
9 |
10 |
|
|
Flat cassette with rare earth intensifying screens
|
•
|
||
|
CCD-type electronic x-ray detector with CsI, high resolution scintillator
|
•
|
||
|
Memory card slot (CompactFlash®)
|
•
|
||
|
Connection to computer via high-speed USB 2.0 port
|
•
|
||
|
Pedestal mount
|
Optional
|
Optional | |
|
Storage compartments
|
•
|
•
|
|
|
Programs:
|
|||
|
Adult/Child Panoramic
|
•
|
•
|
|
|
Adult/Child Hemi Panoramic Right
|
•
|
•
|
|
|
Adult/Child Hemi Panoramic Left
|
•
|
•
|
|
|
Adult/Child TMJ Open/Close Mouth
|
•
|
Optional
|
|
|
Adult/Child Bi-Axial TMJ
|
•
|
Optional
|
|
|
Adult/Child Maxillary Sinus P-A
|
•
|
Optional
|
|
|
Maxillary Sinus L-L
|
•
|
Optional
|
|
|
Improved Orthogonality Dentition
|
•
|
Optional | |
|
Implantologic
|
Optional | ||
|
Technical Data
|
|||
|
Exposure times
|
Panoramic |
15 s adult, 13.5 s child
|
|
| TMJ open/closed mouth |
4 x 2.65 s
|
||
| Cephalometric | 0.2-3 s | ||
| High voltage | 50-80 kVp, in 2 kV steps | ||
| Anode current | 4-12 mA, in 1 mA steps | ||
| Focal spot | 0.5 mm (IEC336) | ||
| Cassette size | Panoramic, implant, sinus, other exams | N/A | 6 x 12 (15 x 30 cm) flat |
| Cephalometric | N/A | 8 x 10 (20.3 x 25.4 cm) | |
| Gray levels | 4096 (12 bits) | N/A | |
| Height of irradiated area on sensor | 5 5/8 (143 mm) | N/A | |
| Required operating system | Windows XP | N/A | |
| Power supply voltage | 120/230 VAC ±10%, single phase, 50/60 Hz | ||
| Current load | 15 A @108 VAC, 8 A @ 230 VAC | ||
| Power rating | 2 kVA | ||
| Weight | w/ceph arm: 402 lbs. (150 kg) w/o ceph arm: 362 lbs. (135 kg) |
||