Dental Panoramic and Cephalometric Systems

Specifications

Rotograph Plus Rotograph Digital
Exposure TimesPanoramic
17 s adult, 14 s child
TMJ
10.4 s adult, 9.4 s child
Cephalometric
0.2-3 s in 19 steps
N/A
Exposure interval
4 min. (1:16 duty cycle)
High voltage
60-85 kVp, in 5 kV steps
Anode current
10 mA
Focal spot
0.5 mm (IEC336)
Power supply voltage
120 VAC ±10%, single phase, 60 Hz
Current load
10 A max
Power rating
1.15 kVA
Cassette sizePanoramic, TMJ
5 x 12 (12.7 x 30 cm), curved
N/A
Cephalometric
8 x 10 (20.3 x 25.4 cm), flat
N/A
Height if irradiated area on sensor
N/A
5 5/8 (143 mm)
Average image magnification
N/A
1.2:1
Image matrix
N/A
1536 x 2805 pixels in binning modality
CCD sensor resolution
N/A
10.4 lp/mm (pixel size: 48µm)
Gray levels
N/A
4096 (12 bits)
Required operating system
N/A
Windows XP/2000
PC connection
N/A
USB 2.0
Cooling
Ambient
Leakage radiation at 1 m
<0.25 mGy/h
Degree of protection
IP 20
Maximum operating temperature range
+50 °F – +104 °F (+10 °C – +40 °C)
Weight
w/ceph arm: 573 lbs. (260 kg) w/o ceph arm: 520 lbs. (236 kg)
540 lbs. (245 kg)

Strato Specifications

Strato Digital
Strato 2000 Film
Essential Features
Motorized column
3-line laser patient positioning
DC high voltage supply (high frequency power converter)
Microcomputer-controlled movements for multiple projection programs

9

10

Flat cassette with rare earth intensifying screens
CCD-type electronic x-ray detector with CsI, high resolution scintillator
Memory card slot (CompactFlash®)
Connection to computer via high-speed USB 2.0 port
Pedestal mount
Optional
Optional
Storage compartments
Programs:
Adult/Child Panoramic
Adult/Child Hemi Panoramic Right
Adult/Child Hemi Panoramic Left
Adult/Child TMJ Open/Close Mouth
Optional
Adult/Child Bi-Axial TMJ
Optional
Adult/Child Maxillary Sinus P-A
Optional
Maxillary Sinus L-L
Optional
Improved Orthogonality Dentition
Optional
Implantologic
Optional
Technical Data
Exposure times
Panoramic
15 s adult, 13.5 s child
TMJ open/closed mouth
4 x 2.65 s
Cephalometric0.2-3 s
High voltage50-80 kVp, in 2 kV steps
Anode current4-12 mA, in 1 mA steps
Focal spot0.5 mm (IEC336)
Cassette sizePanoramic, implant, sinus, other examsN/A6 x 12 (15 x 30 cm) flat
CephalometricN/A8 x 10 (20.3 x 25.4 cm)
Gray levels4096 (12 bits)N/A
Height of irradiated area on sensor5 5/8 (143 mm)N/A
Required operating systemWindows XPN/A
Power supply voltage120/230 VAC ±10%, single phase, 50/60 Hz
Current load15 A @108 VAC, 8 A @ 230 VAC
Power rating2 kVA
Weightw/ceph arm: 402 lbs. (150 kg) w/o ceph arm: 362 lbs. (135 kg)